Technical Specifications
Technical specifications of the Openventi Artificial Ventilator.
This information is under review and will be updated.
Actual version: 1.1-140421
A/CMV Mode: | Volume control |
PEEP: | 0 to 35 cmH2O |
Tidal Volume: | 200 – 2000mL |
Inspiratory pressure: | From 0 to 60 cmH2O |
High Flow Mode | 0 to 60 lt per minute |
CPAP mode: | Yes. Continuous positive nasal pressure |
CPAP pressure | 0-30 cmH20 |
Sound | Yes. Buzzer |
Visuals | Yes. Configurable from LCD or tablet, for pressure and flow |
Variables by alarm | FiO2, pressure, volume, PEEP, obstruction, RR high, disconnection, apnea |
System alarms | Turbine current consumption monitoring for early fault detection |
Type | By pressure |
Range | 5 – 20 cmH2O |
Microcontroller | ESP32 dual core 240MHz |
Operating system | Real Time FreeRTOS |
WiFi | Yes |
Bluetooth | Yes |
Oxygen sensor | Yes |
Measuring range | 0-100% |
Oxygen Mix | At low pressure in mixing chamber |
Technology | Blower / turbine |
Maximum blower pressure | 6 kPa |
Maximum RPM | 45,000 |
Flow measurement | Yes, SENSIRION high precision sensors |
Pressure measurement | Yes, SENSIRION high precision sensors |
Flow measurement on inspiration | Yes |
Flow measurement on expiration | Yes |
Microcontroller | ESP32 dual core 240MHz |
Operating system | Real Time FreeRTOS |
WiFi | Yes |
Bluetooth | Yes |
USB | Micro USB port inside the equipment for updates |
Storage | 8GB microDB card |
Blower Current Monitoring | Yes. Current sensor up to 50A |
Dimensions | 5 inches diagonal |
Touchscreen | Yes |
Touch technology | Resistive |
Screen lock | Yes |
Independent Processor | Yes, microcontroller with ARM architecture, exclusively in charge of drawing the interface |
Waveforms | Pressure, Volume, Flow |
Parameter Configuration | PEEP, Tidal Volume, Respiration rate |
Alarms | Yes |
Patient’s data | Yes |
Input voltage | 100-240 VAC 50/60Hz |
Maximum power consumption | 120W |
Batteries | No |
Material | Stainless steel |
Thickness | 1.2mm |
Machined | Laser cut |